Model | GE Datex Ohmeda Care station ADU S/5 | Ge Datex Ohmeda Aestiva 5 |
Patient Type | Adult / Pediatric | Adult / Pediatric (Optional) |
Tidal Volume | 20 – 1400 ml | 300-1500 ml
50- 300 ml (Optional) |
Flow Meter | Entirely electronically controlled; no needle valves or glass flow tubes. | Glass filled Hg. |
Integrated Ventilator | Microprocessor based, electronically controlled, pneumatically driven ventilator | 7900 Smart Ventilator |
Volume Controlled Ventilation | VCV , PCV , SIMV
Pressure Support (PSVPro) with Apnea Backup ventilation — (optional) |
VCV , PCV , SIMV(Optional)
Pressure Support (PSVPro) with Apnea Backup ventilation — (optional) |
Soda Lime Canister | 0.5 KG Circle Absorber System- Disposable .
Upgraded to higher Volume |
1.4 KG Circle Absorber System |
Monitor Display | 10.4’or 12.4’ | 3’ |
Vaporizer Position | 01 Std. 03 Back up. | 02 or 03 Std. |
Vaporizer Type | Electronic Control Service free, lightweight Aladdin Cassettes | Variable-bypass |
GAS Connections | O2, N20, AIR | O2, N20, AIR |
Special Features | Cardiac bypass case mode | |
Oxygen Measurement | Paramagnetic O2 Sensor Board | Galvanic fuel cell
Life :Approximately 18 months (dependent on usage) |
I:E Ratio (REGULATORS) | 4:1 to 1:4 | 2:1 to 1:8 (increments of 0.5) |
FDA CLEARANCE (Interface Compensation) | Yes | Yes |
CE MARK (MDD)
(Interference compensation) |
Yes | Yes |
We are giving you options. Choose as per your need!